Remote Control
R&S
®
FSPN
448User Manual 1179.4363.02 ─ 01
Manual operation: See " Shift y " on page 181
CALCulate<n>:LIMit<li>:UPPer:SPACing <InterpolType>
This command selects linear or logarithmic interpolation for the calculation of an upper
limit line from one horizontal point to the next.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<InterpolType> LINear | LOGarithmic
*RST: LIN
Manual operation: See " Y-Axis " on page 180
11.7.5.3 Reading out the Results of a Limit Check
CALCulate<n>:LIMit<li>:CLEar[:IMMediate]......................................................................448
CALCulate<n>:LIMit<li>:FAIL?........................................................................................ 448
CALCulate<n>:LIMit<li>:CLEar[:IMMediate]
This command deletes the result of the current limit check.
The command works on all limit lines in all measurement windows at the same time.
Suffix:
<n>
.
Window
<li> irrelevant
Example:
CALC:LIM:CLE
Deletes the result of the limit check.
CALCulate<n>:LIMit<li>:FAIL?
This command queries the result of a limit check in the specified window.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single measurement mode.
Suffix:
<n>
.
Window
<li> Limit line
Return values:
<Result> 0
PASS
Common Analysis and Display Functions