CALCulate<n>:LIMit<li>:LOWer[:DATA] <LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<LimitLinePoints> Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with CALCulate<n>:
LIMit<li>:CONTrol[:DATA]. If not, the R&S FSPN either
adds missing values or ignores surplus values.
*RST: Limit line state is OFF
Default unit: DBM
Manual operation: See " Data Points " on page 181
CALCulate<n>:LIMit<li>:LOWer:MODE <Mode>
This command selects the vertical limit line scaling.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Mode> ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST: ABSolute
Manual operation: See " X-Axis " on page 180
CALCulate<n>:LIMit<li>:LOWer:SHIFt <Distance>
This command moves a complete lower limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
Window
<li> Limit line