Remote Control
R&S
®
FSPN
484User Manual 1179.4363.02 ─ 01
Example:
CAL:RES?
would return, e.g.
Total Calibration Status:
PASSED, Date (dd/mm/yyyy): 12/07/2004,
Time: 16:24:54,Runtime: 00.06
Usage: Query only
Manual operation: See " Alignment Results: " on page 214
DIAGnostic:SERVice:INPut:MC[:DISTance] <Bandwidth>
This command selects the distance of the peaks of the microwave calibration signal for
calibration of the YIG filter.
Parameters:
<Bandwidth> WIDE | SMALl
SMALl
Small offset of combline frequencies.
WIDE
Wide offset of combline frequencies.
Manual operation: See " Calibration Frequency MW " on page 239
DIAGnostic:SERVice:INPut:PULSed:CFRequency <Frequency>
This command defines the frequency of the calibration signal.
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic:SERVice:INPut[:SELect] on page 485.
Manual operation:
See " Calibration Frequency RF " on page 239
DIAGnostic:SERVice:INPut:RF[:SPECtrum] <Bandwidth>
This command selects the bandwidth of the calibration signal.
Parameters:
<Bandwidth> NARRowband | BROadband
NARRowband
Narrowband signal for power calibration of the frontend.
BROadband
Broadband signal for calibration of the IF filter.
Manual operation: See " Spectrum " on page 239