cobas p 512 Part E - Troubleshooting
Operator's Manual - Version 1.6 - 10/2015 5-11
Return code Error Reason
-305
Plausibility check
failed. Falsified result
of liquid level
detection suspected.
Serum's upper edge detected just below
the label's lower edge, but ...
... due to a narrow viewport beside the
label, the serum's upper edge is suspected
to really be above the label's lower edge.
Example: