Part D - Maintenance cobas p 512
4-24 Operator's Manual - Version 1.6 - 09/2015
4.7. Replacing parts
4.7.1 Replacing the mains fuse
The mains fuse is located above the mains connection at the rear of the system.
WARNING
Electric Shock!
Do not replace the fuse until you have unplugged the power cable!
Figure 4-28: Interface block
1. Exit the control software via the [Logout] button in the upper right-hand corner of the
monitor. A pop-up window is displayed.
2. Touch the [Logout] button in this pop-up window to confirm the Logout procedure.
The program will be shut down and the Login screen is displayed.
3. To shut down the system, touch the [Shutdown] button in the lower left-hand corner of
the monitor.
4. When requested, turn the main switch to the position ‘O’ to turn off the system.
5. Remove the mains supply cable from the mains supply connection on the system.