Part B - System description cobas p 512
2-8 Operator's Manual - Version 1.6 - 10/2015
2.2.1 Schematic diagram
Figure 2-2: Schematic diagram of the cobas p 512 system
2.2.2 System safety devices
Areas that present a hazard during operations are secured by a protective door. The
operating area of the movable sorting unit is secured with protective doors on the input and
output sorters.
The system can only be started when the protective doors are closed. The protective door is
locked at the front by pressing the [Run] button and can only be opened when in the standby
mode or powered off.