Part D - Maintenance cobas p 512
4-8 Operator's Manual - Version 1.6 - 09/2015
4.4. Weekly device care
CAUTION
Switch off the device before cleaning! Always wear protective clothing when
cleaning! There is a risk of infection!
4.4.1 Cleaning the decapper module
The decapper module is fastened to the housing so that it can be swung out or also reached
through a door.
4.4.1.1 Cleaning the cap waste system and sensors
1. Switch the system off.
2. Disconnect the compressed air supply.
3. Loosen the screw on the door located at the rear of the decapper module.
4. Open the door of the decapper module.
5. Remove the waste tube.
Figure
4-1: Screw on the door of the
decapper module
Figure
4-2: Waste tube removal