Operator’s Manual · Version 3.0 139
Multiplate® analyzer 5 Operation
Multiplate system control
3
In the instruction window, choose the Start electronic QC button.
When the electronic control measurement is complete a window is displayed that
contains the results.
Figure 5-50 Results of electronic control measurement
4
Check whether the results are within the ranges.
q
If results fail to pass the electronic control criteria, the first step is to repeat the
procedure. If the results continue to fail, contact your Roche Service representative for
further advice.
If any of the channels fail to pass the control measurement due to a short circuit
or a bad electronic contact a second message informs you of the channels that are
outside the reference range.