Operator’s Manual · Version 3.0 171
Multiplate® analyzer 7 Maintenance
Maintenance and cleaning of the electronic pipette
6
Close the cover of the battery compartment.
Figure 7-4 Closing battery compartment
After replacing the battery, a reference run becomes automatically due.
7
Perform a reference run.
u For details on performing a reference run, see Performing a reference run (p. 180).
When the reference run is finished, the display automatically returns to the
normal view.
Figure 7-5 Normal display view
8
When the reference run is complete, plug the pipette cable into the electronic
pipette.
Figure 7-6 Connecting the pipette cable