Roche Diagnostics
Operator’s Manual · Version 3.0 181
Multiplate® analyzer 7 Maintenance
Maintenance and cleaning of the electronic pipette
Figure 7-25 Unplugging the pipette
2
Simultaneously press the MENU and ENTER buttons.
Figure 7-26 Initiating reference run mode
On screen, the rEF text indicates that the pipette is in reference run mode.
Figure 7-27 Reference run mode, display