Signal generator settings
R&S
®
SGS100A
129User Manual 1173.9105.02 ─ 13
"Pulse Video Out"
Output of the internally generated pulse video (modulating) signal.
The video signal level corresponds to the RF envelope.
"Pulse Mod Ext Trigger"
Input for an external trigger signal, used to trigger the pulse genera-
tor.
"Pulse Mod Ext Source"
Input for an externally provided pulse modulation signal.
Used when an external pulse modulator source is provided at the
connector.
Remote command:
:CONNector:TRIGger:OMODe on page 230
Trigger Mode
Selects between continuous pulse modulation or pulse modulation triggered by an
external signal.
"Auto"
Generates the modulation signal continuously.
"Ext Single"
Generates the signal each time an external trigger event occurs.
"Ext Gated"
Generates the signal triggered by an external gate signal.
Remote command:
[:SOURce<hw>]:PULM:TRIGger:MODE on page 281
Ext. Trigger Input Slope
Available only for "Trigger Mode > Ext Triggered "
Sets the polarity of the active slope of an applied external trigger signal.
Remote command:
[:SOURce<hw>]:PULM:TRIGger:EXTernal:SLOPe on page 281
Gate Input Polarity
Available only for "Trigger Mode > Ext Gated "
Sets the polarity of the active slope of an applied gate signal.
Remote command:
[:SOURce<hw>]:PULM:TRIGger:EXTernal:GATE:POLarity on page 281
9.7 Trigger connector settings
The dialog provides settings to determine the signal at the input/output of the multi-pur-
pose "TRIG" connector.