Understanding the R&S SGMA-GUI software
R&S
®
SGS100A
89User Manual 1173.9105.02 ─ 13
Use the IP address to optimize scanning for instruments within a subnet.
Parameters:
<Address> string
Example: See Example "Scanning a subnet for SGMA instruments"
on page 84.
Manual operation: See "IP Address" on page 64
:INSTruments:SCAN:SNET:PLENgth <Number>
Sets the prefix length of the subnet mask.
Use the prefix length to optimize scanning for instruments within a subnet. A higher
value accelerates the scanning procedure but lowers the ability to detect all instru-
ments within the subnet.
Parameters:
<Number> integer
Range: 18 to 30
*RST: 18
Default unit: bit
Example: See Example "Scanning a subnet for SGMA instruments"
on page 84.
Manual operation: See "Prefix Length" on page 64
:INSTruments:SCAN:SNET[:STATe] <State>
Activates scanning of a subnet with given IP address and prefix length of the subnet.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: n.a. (no preset. default: OFF)
Example: See Example "Scanning a subnet for SGMA instruments"
on page 84.
Manual operation: See "Scan Subnet" on page 64
:INSTruments:SERial <Serial>
Enters the serial number as instrument's identification while using the USB interface for
remote control.
Parameters:
<Serial> <SerialNumberInstr#1>, <SerialNumberInstr#2>,...
Example: SeeExample "Configuring SGMA instruments" on page 85.
Manual operation: See "Serial Number" on page 67