Signal generator settings
R&S
®
SGS100A
121User Manual 1173.9105.02 ─ 13
Source...........................................................................................................122
Frequency.....................................................................................................122
Level Offset State,Level Offset..................................................................... 122
Filter..............................................................................................................122
Filter Length..................................................................................................123
Auto Once.....................................................................................................123
Noise Content............................................................................................... 123
Timeout......................................................................................................... 123
Default Aperture Time...................................................................................123
Aperture Time............................................................................................... 124
S-Parameter..................................................................................................124
Enable Logging.............................................................................................124
Sensor type and serial number
Indicates the type and the serial number of the selected R&S NRP power sensor, and
the channel the sensor is assigned to.
The displayed Level (Peak) / Level (Average) values correspond to the particular sen-
sor.
Remote command:
:SENSe<ch>[:POWer]:TYPE? on page 258
:SENSe<ch>[:POWer]:SNUMber? on page 257
Level (Peak) / Level (Average)
Indicates the measured peak or average level value.
You can also change the unit for the results display: Watt, dBm or dBμV.
Note: Peak level measurements are provided if the power sensor supports this feature.
Remote command:
:READ<ch>[:POWer]? on page 251
:SENSe<ch>:UNIT[:POWer] on page 251
Sensor Mapping
Accesses the NRP sensor mapping dialog.
Sensor Settings
One tab per sensor provides the corresponding setting parameters.
State ← Sensor Settings
Activates level measurement.
Remote command:
:INITiate<hw>[:POWer]:CONTinuous on page 250
To query the availability of a sensor at a given connector, use the command :
SENSe<ch>[:POWer]:STATus[:DEVice]? on page 258.
Zero ← Sensor Settings
Activates the auto zeroing.
For details, see "About zeroing" on page 118.
Level and power-on settings