Remote control commands
R&S
®
SGS100A
226User Manual 1173.9105.02 ─ 13
Setting parameters:
<UnlockId> Number
Unlock ID which uniquely identifies the controller to the instru-
ment. The value must match the Controller ID <Lock Request
Id> set with the command :LOCK?.
0
Clear lock regardless of locking state
Usage: Setting only
Manual operation: See "Exclusive Access" on page 66
13.4 Preset commands
The preset commands are not bundled into one subsystem. Therefore, they are listed
separately in this section. In addition, a specific preset command is provided for each
digital standard and for the fader. These specific commands are described in the asso-
ciated subsystems.
The following presetting actions are available:
●
Activating the default state of all internal instrument functions (*RST on page 223).
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. TCP/IP address or reference oscillator source settings.
●
Activating the original state of delivery (factory reset, :SYSTem:FPReset
on page 226). Only functions that are protected by a password remain unchanged
and the passwords themselves.
:SOURce<hw>:PRESet
:SYSTem:PRESet
Triggers an instrument reset. It has the same effect as:
●
The *RST command
●
The "SGMA-GUI > Instrument Name > Preset" function.
However, the command does not close open GUI dialogs like the function does.
For an overview of the settings affected by the preset function, see Chapter 9.9, "Pre-
set", on page 135.
Example:
SYST:PRES
All instrument settings (also the settings that are not currently
active) are reset to their default values.
Usage: Setting only
:SYSTem:FPReset
Triggers an instrument reset to the original state of delivery.
Preset commands