Signal generator settings
R&S
®
SGS100A
92User Manual 1173.9105.02 ─ 13
Remote command:
[:SOURce]:OPMode on page 259
9.2 Frequency/Phase settings
Depending on the installed options, the instrument provides an adjustable output fre-
quency in the frequency range of 1 MHz to 6 GHz or to 12.75 GHz.
Access:
Select "SGMA-GUI > Instrument Name > Frequency/Phase".
The frequency and phase-related settings are provided in this dialog.
Configuring the RF frequency
1. To change the RF frequency of the selected instrument, perform one of the follow-
ing:
a) Select "SGMA-GUI > Instrument Name > Frequency/Phase > Frequency"
b) Select "SGMA-GUI main panel > Freq".
The value includes a possible frequency offset.
2. To enable a frequency offset, for example, to consider the frequency shift of a
downstream instrument, set the parameter "Frequency/Phase > Offset".
Changes of the RF frequency have an immediate effect on the output signal.
Frequency
Sets the RF frequency at the RF output connector of the selected instrument.
The displayed value does not consider an enabled frequency offset (Offset).
Remote command:
See [:SOURce]:FREQuency[:CW|FIXed] on page 259
Offset
Sets a frequency offset.