Preparing for use
R&S
®
SGS100A
21User Manual 1173.9105.02 ─ 13
Left = Stacked correctly, same dimensions
Middle = Stacked correctly, different dimensions
Right = Stacked incorrectly, too many products
3.
NOTICE! Overheating can damage the product.
Prevent overheating as follows:
● Keep a minimum distance of 10 cm between the fan openings of the product
and any object in the vicinity.
● Do not place the product next to heat-generating equipment such as radiators
or other products.
3.4.2 Mounting the R&S SGS in a rack
To prepare the rack
1. Observe the requirements and instructions in "Setting up the product" on page 12.
2.
NOTICE! Insufficient airflow can cause overheating and damage the product.
Design and implement an efficient ventilation concept for the rack.
To mount the R&S SGS in a rack
1. Use an adapter kit that fits the dimensions of the R&S SGS to prepare the instru-
ment for rack mounting.
a) Order the rack adapter kit designed for the R&S SGS. For the order number,
see data sheet.
b) Mount the adapter kit. Follow the assembly instructions provided with the
adapter kit.
2. Push the R&S SGS onto the shelf until the rack brackets fit closely to the rack.
3. Tighten all screws at the rack brackets with a tightening torque of 1.2 Nm to secure
the R&S SGS in the rack.
To unmount the R&S SGS from a rack
1. Loosen the screws at the rack brackets.
2. Remove the R&S SGS from the rack.
3. If placing the R&S SGS on a bench top again, unmount the adapter kit from the
R&S SGS. Follow the instructions provided with the adapter kit.
Setting up the R&S
SGS