Remote control commands
R&S
®
SGS100A
258User Manual 1173.9105.02 ─ 13
Note: When measuring the RF signal, the sensor considers the corresponding correc-
tion factor at that frequency, and uses the level setting of the instrument as reference
level.
Parameters:
<Source> A | USER | RF
*RST: A
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 247.
Manual operation: See "Source" on page 122
:SENSe<ch>[:POWer]:STATus[:DEVice]?
Queries if a sensor is connected to the instrument.
Return values:
<Status> 0 | 1 | OFF | ON
*RST: 0
Example:
:SENS1:STAT?
Response: 1
A sensor is connected.
Usage: Query only
Manual operation: See "State" on page 121
:SENSe<ch>[:POWer]:SVERsion?
Queries the software version of the connected R&S NRP power sensor.
Return values:
<SVersion> string
Example:
:SENS1:POW:SVER?
Queries the software version of the power sensor.
Usage: Query only
:SENSe<ch>[:POWer]:TYPE?
Queries the sensor type. The type is automatically detected.
Return values:
<Type> string
Example:
:SENS1:TYPE?
Queries the type of sensor.
Response: NRP-Z21
The R&S NRP-Z21 sensor is used.
Usage: Query only
SENSe, READ, INITiate and SLISt subsystems