EasyManua.ls Logo

Rohde & Schwarz R&S SGS100A - [:Source]:Power:scharacteristic

Rohde & Schwarz R&S SGS100A
326 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SGS100A
275User Manual 1173.9105.02 ─ 13
This value does not consider a specified offset. The command [:SOURce]:POWer[:
LEVel][:IMMediate][:AMPLitude] sets the level of the "Level" display, that
means the level containing offset.
Parameters:
<Amplitude> float
Range: -20 to 25
Increment: 0.01
*RST: -10
Example:
POW:POW 15
sets the RF level at output to 15 dBm.
Manual operation: See "Level/Level Offset" on page 74
[:SOURce]:POWer:SCHaracteristic <Characteristic>
Selects the characteristic for the level setting.
Parameters:
<Characteristic> AUTO | UNINterrupted | CVSWr | USER | MONotone
UNINterrupted
uninterrupted level setting
CVSWr
constant-VSWR
MONotone
strictly monotone
*RST: AUTO
Example: See Chapter 13.1.4, "Advanced tasks to optimize performance",
on page 218.
Manual operation: See "Setting Characteristic" on page 100
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude] <Amplitude>
Sets the RF level at the RF output connector of the instrument.
Parameters:
<Amplitude> float
Range: -20 to 25
Increment: 0.01
*RST: -10
Example: See Chapter 13.1.4, "Advanced tasks to optimize performance",
on page 218.
Manual operation: See "Level/Level Offset" on page 74
SOURce:POWer subsystem

Table of Contents

Related product manuals