Remote control commands
R&S
®
SGS100A
275User Manual 1173.9105.02 ─ 13
This value does not consider a specified offset. The command [:SOURce]:POWer[:
LEVel][:IMMediate][:AMPLitude] sets the level of the "Level" display, that
means the level containing offset.
Parameters:
<Amplitude> float
Range: -20 to 25
Increment: 0.01
*RST: -10
Example:
POW:POW 15
sets the RF level at output to 15 dBm.
Manual operation: See "Level/Level Offset" on page 74
[:SOURce]:POWer:SCHaracteristic <Characteristic>
Selects the characteristic for the level setting.
Parameters:
<Characteristic> AUTO | UNINterrupted | CVSWr | USER | MONotone
UNINterrupted
uninterrupted level setting
CVSWr
constant-VSWR
MONotone
strictly monotone
*RST: AUTO
Example: See Chapter 13.1.4, "Advanced tasks to optimize performance",
on page 218.
Manual operation: See "Setting Characteristic" on page 100
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude] <Amplitude>
Sets the RF level at the RF output connector of the instrument.
Parameters:
<Amplitude> float
Range: -20 to 25
Increment: 0.01
*RST: -10
Example: See Chapter 13.1.4, "Advanced tasks to optimize performance",
on page 218.
Manual operation: See "Level/Level Offset" on page 74
SOURce:POWer subsystem