Remote control commands
R&S
®
SGS100A
282User Manual 1173.9105.02 ─ 13
EXTernal
The pulse modulation is triggered by an external trigger event.
The trigger signal is supplied via the trigger connector.
EGATe
The pulse modulation is gated by an external gate signal. The
trigger signal is supplied via the trigger connector.
*RST: AUTO
Example:
PULM:TRIG:MODE EXT
selects triggering by an external trigger event.
Manual operation: See "Trigger Mode" on page 129
[:SOURce<hw>]:PULM:WIDTh <Width>
Sets the width of the generated pulse. The width determines the pulse length. The
pulse width must be at least 20ns less than the set pulse period.
Parameters:
<Width> float
Range: 20 ns to 100 s
Increment: 10 ns
*RST: 2 us
Example:
PULM:WIDT 33 us
sets a width of 33 us for the pulse.
Manual operation: See "Pulse Width" on page 127
13.19 SOURce:ROSCillator subsystem
[:SOURce]:ROSCillator:EXTernal:FREQuency.................................................................. 282
[:SOURce]:ROSCillator:OUTPut:FREQuency....................................................................283
[:SOURce]:ROSCillator:SOURce..................................................................................... 283
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue.............................................................. 283
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe].............................................................284
[:SOURce]:ROSCillator:EXTernal:SBANdwidth..................................................................284
[:SOURce]:ROSCillator:EXTernal:FREQuency <ExtFreq>
Selects the frequency of the external reference.
Parameters:
<ExtFreq> 10MHZ | 100MHZ | 1000MHZ | 13MHZ
13MHZ requires RF board with part number 1419.5308.02.
To find out the RF board installed in the instrument:
Select "SGMA-GUI > instrument name > Setup > Hardware
Config" > "RF Assembly"
Observe the part number of the assembly "RfBoard".