Signal generator settings
R&S
®
SGS100A
127User Manual 1173.9105.02 ─ 13
Remote command:
[:SOURce<hw>]:PULM:TRIGger:EXTernal:IMPedance on page 281
9.6.2 Pulse generator settings
Access:
Select "SGMA-GUI > Instrument Name > Pulse Modulation > Pulse Genrator".
Comprises the settings necessary to configure the internal pulse modulation signal.
Pulse Mode
Sets the operating mode of the pulse generator.
"Single"
Generates a single pulse in one pulse period, see Figure 9-4.
"Double"
Generates two pulses in one pulse period, see Figure 9-5.
Remote command:
[:SOURce<hw>]:PULM:MODE on page 279
Pulse Period
Sets the repetition rate of the generated pulse signal, see "Pulse modulation signal
waveforms" on page 125.
Remote command:
[:SOURce<hw>]:PULM:PERiod on page 279
Pulse Width
Sets the pulse duration of the generated pulse signal, see "Pulse modulation signal
waveforms" on page 125.
Remote command:
[:SOURce<hw>]:PULM:WIDTh on page 282
Double Pulse Width
Sets the width of the second pulse, see Figure 9-5.
Pulse modulation