General instrument settings and instrument setup
R&S
®
SGS100A
155User Manual 1173.9105.02 ─ 13
Note: For background information and description of the syntax of the VISA resource
strings, refer to the description of the corresponding interface in Chapter 12.1, "Remote
control interfaces and protocols", on page 175.
Remote command:
:SYSTem:COMMunicate:HISLip:RESource? on page 295
:SYSTem:COMMunicate:NETWork:RESource? on page 294
:SYSTem:COMMunicate:SOCKet:RESource? on page 296
:SYSTem:COMMunicate:USB:RESource? on page 296
:SYSTem:COMMunicate:PCIexpress:RESource? on page 296
Device Identity
Triggers the device identification function. The [LAN] LED on the front panel of the
selected instrument blinks.
See also Chapter 8.4.4, "Bidirectional instrument identification", on page 79.
10.9 Factory preset
The "Factory Preset" dialog provides a function to reset the instrument's settings to
their factory states.
Access:
1. Select "SGMA-GUI > Instrument Name > Setup > Factory Preset".
2. Select "Execute Factory Preset".
The instrument's settings are reset to their factory states. Security settings and
user data are not affected.
Settings
Execute Factory Preset...............................................................................................155
Execute Factory Preset
Reset the instrument's settings to their factory state.
Note: Because "Factory Preset" resets the "Remote Channel Settings" and "Network
Settings" to the default values, executing factory preset via remote control can termi-
nate the connection to the instrument, if these settings had been configured to values
different to the default ones.
Factory preset