General instrument settings and instrument setup
R&S
®
SGS100A
156User Manual 1173.9105.02 ─ 13
The factory preset function resets nearly all instrument settings. In addition to the regu-
lar preset, a "Factory Preset" resets also the following values:
Power on settings ("Level" dialog)
Network settings including hostname ("Setup > Network Setting" dialog)
Remote Channel settings ("Setup > Remote Channel" dialog)
Eco mode state ("Setup > Eco Mode" dialog)
To maintain security, password settings and all settings protected by these passwords
like disabled USB and LAN connections are not changed.
Not affected by the "Factory Preset" are also user data, lists or instrument settings
files, created, for example, with the "File Save As" function.
Remote command:
:SYSTem:FPReset on page 226
10.10 Eco mode
This energy-saving mode is available only for instruments equipped with
optionR&S SGS-B112/B112V.
Access:
Select "SGMA-GUI > Instrument Name > Setup > Eco Mode".
With enabled "Eco Mode 1", the doubler stage in a 12.75 GHz instrument is perma-
nently switched off to reduce power consumption and the maximum frequency is
limited to 6 GHZ. An enabled "Eco Mode" is indicated by a green coloring of the
frequency range in the R&S SGMA-GUI.
The state of this parameter is not affected by an instrument "Preset". This parame-
ter is influenced only by the Factory preset.
Running internal adjustments in eco mode
See "Running internal adjustments in eco mode" on page 142.