Remote control commands
R&S
®
SGS100A
284User Manual 1173.9105.02 ─ 13
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe] <State>
Determines whether the calibrated (OFF) or a user-defined (ON) adjustment value is
used for fine adjustment of the frequency.
If user-defined values are used, the instrument is no longer in the calibrated state.
However, the calibration value is not changed and the instrument resumes the calibra-
ted state after sending the com-
mand :SOURce:ROSCillator:INTernal:ADJust:STATe OFF.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example:
ROSC:SOUR INT
Selects the internal source.
ROSC:ADJ ON
Activates use of a user-defined adjustment value.
ROSC:ADJ:VAL 1400
Sets the adjustment value to 1400.
Manual operation: See "Adjustment Active" on page 98
[:SOURce]:ROSCillator:EXTernal:SBANdwidth <SBandwidth>
Sets the synchronization bandwidth for an external reference signal.
Parameters:
<SBandwidth> WIDE | NARRow
NARRow
The synchronization bandwidth is 40 Hz.
WIDE
Synchronization bandwidth is 250 Hz.
*RST: n.a. (factory preset)
Example:
ROSC:SOUR EXT
Selects the external source.
ROSC:EXT:FREQ 10 MHz
Informs the instrument that the external reference has a fre-
quency of 10 MHz.
ROSC:EXT:SBAN WID
Selects wideband setting for synchronization bandwidth.
Manual operation: See "Synchronization Bandwidth" on page 97
13.20 STATus subsystem
This system contains the commands for the status reporting system. See also Chap-
ter 12.5, "Status reporting system", on page 193 for detailed information.
STATus subsystem