Performing configuration tasks
R&S
®
SGS100A
172User Manual 1173.9105.02 ─ 13
Restoring the default settings
Use the "R&S SGMA-GUI main panel > Instrument name > Preset" or the "R&S
SGMA-GUI main panel > Instrument name > Setup > Factory Preset" function to return
the instrument to its predefined state.
For information on how to adjust the quality characteristics of the RF output signal, i.e.
to optimize the quality characteristics of RF output signal, refer to Chapter 9.5, "Level
and power-on settings", on page 98.
For information on how to adjust the reference oscillator, e.g. to allow the frequency of
the internal reference oscillator to be impaired, refer to Chapter 9.4, "Reference oscilla-
tor", on page 95.
11.12 How to install a new firmware version on the instru-
ment
You can update the firmware of the R&S SGS.
Firmware installation via R&S SGMA-GUI
Install or update the firmware of the R&S SGS before installing or updating the soft-
ware R&S SGMA-GUI.
1. Select "SGMA-GUI main panel > Instrument Name > Setup > Maintenance > Oper-
ation > Install firmware package".
2. Press "Select Package" and navigate to the directory of the new firmware.
3. If several instruments require a firmware update, enable "Update All" to accelerate
the update procedure.
All instruments that are in active state and are connected to this controller are
updated simultaneously.
4. Enter the "Security Password".
5. Confirm the update with "Accept".
The software transfers the firmware file and automatically starts the update proce-
dure. During the update, the message "Updating Firmware" is displayed in the
"Info" line. The update process is indicated by an LED running light.
Note: The update procedure requires a restart of the instrument. The restart is per-
formed automatically. The instrument is not accessible during that time.
6. Wait until the message "Updating Firmware" disappears. After the message disap-
pears, the update is complete.
The [POWER ON/STANDBY] LED is green.
How to install a new firmware version on the instrument