Performing configuration tasks
R&S
®
SGS100A
173User Manual 1173.9105.02 ─ 13
Tip: Calibration error. If the "Info" line shows the message "Calibration Error",
select "SGMA-GUI main panel > Instrument Name > Setup > Internal Adjustments
> Adjust All" to trigger internal adjustment.
7. If necessary, install the new R&S SGMA-GUI.
For detailed description, refer to Chapter 3.13.1, "Installing the R&S SGMA-GUI
software", on page 28
8. If you connect the instrument and the controller/PC over the PCIe interface and the
external PC does not support hot-plugging, restart the external PC.
Firmware update through a session control protocol (SCP)
Install or update the firmware of the R&S SGS before installing or updating the soft-
ware R&S SGMA-GUI.
1. Connect the R&S SGS and a Windows PC to the same network.
2. On the PC, open a windows explorer window.
3. To connect to the R&S SGS, enter the name of the instrument or its IP address in
the windows taskbar.
4. Enter the user name and password to connect to the R&S SGS. The default user
name is instrument and the password is instrument.
A folder opens, containing the share and the update folder.
5. Open the update folder.
6. Copy the new firmware update file into the folder.
The update starts automatically.
Unsuccessful or erroneous firmware update
An erroneous or unsuccessful installation of firmware update package is indicated by a
combination of one orange and red LEDs on the front panel.
Refer to the service manual for a description of the displayed error code or contact the
customer support center, see Chapter 14.5, "Contacting customer support",
on page 312.