EasyManua.ls Logo

Rohde & Schwarz R&S SGS100A - Table of Contents

Rohde & Schwarz R&S SGS100A
326 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SGS100A
247User Manual 1173.9105.02 ─ 13
Example: Performing a power measurement with a fixed filter
Prerequisite: The sensor is connected to the instrument and mapped to the first sensor
channel.
SENSe1:SOURce RF
// Sensor measures the power of the RF signal.
SENSe1:FILTer:TYPE NSRatio
// Selects fixed noise filter mode.
SENSe1:FILTer:NSRatio 0.02 DB
// Sets the maximum noise component in the result to 0.02 DB.
SENSe1:FILTer:NSRatio:MTIMe 10
//Limits the settling time to 10 seconds.
SENSe1:APERture:DEFault:STATe 0
// Deactivates the default aperture time of the sensor.
SENSe1:APERture:TIMe 10e-6
// Sets the aperture time to 10 us.
SENSe1:UNIT DBM
// Selects unit dBm for the measured value.
INITiate:CONTinuous ON
// Switches the continous power measurement on.
READ?
// Triggers the measurement and displays the results.
:SLISt[:LIST]?................................................................................................................ 248
:SLISt:SCAN[:STATe]......................................................................................................248
:SLISt:SCAN:LSENsor....................................................................................................248
:SLISt:SCAN:USENsor................................................................................................... 249
:SLISt:CLEar:LAN.......................................................................................................... 249
:SLISt:CLEar:USB..........................................................................................................249
:SLISt:CLEar[:ALL].........................................................................................................250
:SLISt:ELEMent<ch>:MAPPing........................................................................................250
:SLISt:SENSor:MAP.......................................................................................................250
:INITiate<hw>[:POWer]:CONTinuous............................................................................... 250
:READ<ch>[:POWer]?....................................................................................................251
:SENSe<ch>:UNIT[:POWer]............................................................................................251
:SENSe<ch>[:POWer]:APERture:DEFault:STATe.............................................................. 252
:SENSe<ch>[:POWer]:APERture:TIMe.............................................................................252
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect.......................................................252
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe........................................................253
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?.........................................................253
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?.................................................................. 253
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER].................................................................. 254
:SENSe<ch>[:POWer]:FILTer:NSRatio............................................................................. 254
SENSe, READ, INITiate and SLISt subsystems

Table of Contents