Remote control commands
R&S
®
SGS100A
293User Manual 1173.9105.02 ─ 13
Manual operation: See "IP Address" on page 153
:SYSTem:COMMunicate:NETWork:IPADdress:MODE <Mode>
Selects manual or automatic setting of the IP address.
Parameters:
<Mode> AUTO | STATic
*RST: n.a. (factory preset: AUTO)
Example:
SYSTem:COMMunicate:NETWork:IPADdress:MODE AUTO
The IP address is assigned automatically (DHCP)
Manual operation: See "Address Mode" on page 153
:SYSTem:COMMunicate:NETWork:MACaddress <MacAddress>
Queries the MAC address of the network adapter.
Parameters:
<MacAddress> string
Example:
SYST:COMM:NETW:MAC
queries the MAC address.
:SYSTem:COMMunicate:NETWork:STATus?
Queries the network configuration state.
Return values:
<State> 0 | 1 | OFF | ON
Usage: Query only
:SYSTem:COMMunicate:NETWork:RESTart
Restarts the network connection to the instrument, terminates the connection and sets
it up again.
Example:
SYSTem:COMMunicate:NETWork:RESTart
Usage: Event
:SYSTem:COMMunicate:NETWork[:COMMon]:HOSTname <Hostname>
Sets the individual host name of the R&S SGS.
Note: it is recommended that you do not change the host name in order to avoid prob-
lems with the networdk connection. However, if you change the host name be sure to
use an unique name.
The host name is a protected parameter, To change it, first disable protection level 1
with command :SYSTem:PROTect<ch>[:STATe] on page 298.