Remote control commands
R&S
®
SGS100A
253User Manual 1173.9105.02 ─ 13
Manual operation: See "S-Parameter" on page 124
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe <State>
Activates the use of the S-parameter correction data.
Note: If you use power sensors with attenuator, the instrument automatically activates
the use of S-parameter data.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
:SENSe1:POWer:CORRection:SPDevice:STATe 1
Activates the use of the S-parameters correction data.
Manual operation: See "S-Parameter" on page 124
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?
Queries the list of the S-parameter data sets that have been loaded to the power sen-
sor.
Return values:
<List> string list
*RST: 0
Usage: Query only
Manual operation: See "S-Parameter" on page 124
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
Queries the current filter length in filter mode AUTO (:SENSe<ch>[:POWer]:
FILTer:TYPE)
Return values:
<Auto> float
Range: 1 to 65536
Example:
:SENS1:FILT:TYPE AUTO
Selects auto filter.
:SENS1:FILT:LENG:AUTO?
Queries the automatically set filter length.
Response: 1024
Usage: Query only
Manual operation: See "Filter Length" on page 123
SENSe, READ, INITiate and SLISt subsystems