Signal generator settings
R&S
®
SGS100A
128User Manual 1173.9105.02 ─ 13
Remote command:
[:SOURce<hw>]:PULM:DOUBle:WIDTh on page 279
Pulse Delay
Sets the pulse delay. The pulse delay determines the time that elapses after a trigger
event before pulse modulation starts, see Figure 9-4.
The pulse delay is not effective for double pulse generation.
Remote command:
[:SOURce<hw>]:PULM:DELay on page 278
Double Pulse Delay
Sets the delay between the start of the first pulse and the start of the second pulse,
see Figure 9-5.
Remote command:
[:SOURce<hw>]:PULM:DOUBle:DELay on page 279
9.6.3 Pulse Connector/Trigger settings
Access:
Select "SGMA-GUI > Instrument Name > Pulse Modulation > Connector/Trigger".
Comprises the settings necessary to configure the signal at the multi-purpose
"TRIG" connector in the external trigger mode.
Trigger Connector Mode
Determines the signal at the input/output of the multi purpose [TRIG] connector.
"Signal Valid"
Output of high signal to mark valid frequency and level settings.
"Not Signal Valid"
Output of high signal to mark the transition state when frequency and
level change.
Pulse modulation