Remote control commands
R&S
®
SGS100A
231User Manual 1173.9105.02 ─ 13
:DIAGnostic[:MEASure]:POINt? <Name>
Triggers voltage or temperature measurement at the specified test point and returns
the measured value.
Use the command :DIAGnostic:POINt:CATalog? to retrieve a list of the available
test points.
For description of the test points, see the service manual.
Query parameters:
<Name> string
Return values:
<Value> number
Default unit: V or °C
Example: See Chapter 13.1.1, "Performing general tasks for instrument
setup", on page 213.
Usage: Query only
13.8 EXTension subsystem
:EXTension:REMote:STATe?........................................................................................... 231
:EXTension:INSTruments:NAME?....................................................................................232
:EXTension:INSTruments:SCAN[:STATe].......................................................................... 232
:EXTension:INSTruments:REMote:CHANnel?................................................................... 232
:EXTension:INSTruments:REMote:LAN:NAME?................................................................ 232
:EXTension:INSTruments:REMote:SERial?.......................................................................233
:EXTension:BUSY[:STATe]?............................................................................................ 233
:EXTension:SELect.........................................................................................................233
:TEST:EXTension:CONNection?...................................................................................... 234
:TEST:EXTension:CONNection:RF?.................................................................................234
:EXTension:REMote:STATe?
Queries the state of the remote control connection to the extension.
Return values:
<State> 0 | 1 | OFF | ON
1|ON
Connected
0|OFF
Not connected
Example: See Chapter 13.1.5, "Enabling and configuring an extension
mode", on page 219.
Usage: Query only
Manual operation: See "Available Instruments" on page 138