Signal generator settings
R&S
®
SGS100A
95User Manual 1173.9105.02 ─ 13
Source
Selects the source of the local oscillator signal.
"Int"
The instrument uses the built-in local oscillator.
"Ext"
The signal fed-in at the "REF/LO IN" input connector is used as signal
source.
An icon in the block diagram indicates that an external LO source is
used.
Note: The local oscillator input/output requires the additional software
option R&S SGS-K90.
Remote command:
[:SOURce]:LOSCillator:SOURce on page 260
REF/LO Output
Determines the signal provided at the output connector [REF/LO OUT].
See also Chapter 11.6, "How to define the signal at the REF/LO OUT connector",
on page 167.
"OFF"
No signal is provided.
"LO"
The signal of the local oscillator (LO) is available at the [REF/LO
OUT] connector.
"REF"
The signal of the reference oscillator is available at the [REF/LO
OUT] connector.
Remote command:
:CONNector:REFLo:OUTPut on page 230
9.4 Reference oscillator
The R&S SGS is equipped with an internal TXCO reference oscillator. If equipped with
the additional HW option -B1, an OXCO can be used as an internal reference fre-
Reference oscillator