Signal generator settings
R&S
®
SGS100A
103User Manual 1173.9105.02 ─ 13
9.5.3 Power-On/EMF
Access:
► Select "SGMA-GUI > Instrument Name > Level > Power-On/EMF".
This dialog comprises the settings for the power-on behavior of the instrument.
Power-On State
Selects the state which the RF output takes after the instrument is switched on.
"RF Off"
The output is deactivated when the instrument is switched on.
"Previous Setting"
When the instrument is switched on, the output takes the same state
as it had when the instrument was switched off.
Remote command:
:OUTPut[:STATe]:PON on page 244
9.5.4 ALC
Access:
► Select "SGMA-GUI > Instrument Name > Level > ALC".
You can use automatic level control (ALC) with almost all applications, especially
I/Q modulation. Deactivate ALC for certain settings in the baseband and when I/Q
impairments are active ("Impairments > On").
Level and power-on settings