Remote control commands
R&S
®
SGS100A
244User Manual 1173.9105.02 ─ 13
Manual operation: See "Level Range" on page 102
:OUTPut:AFIXed:RANGe:UPPer?
Queries the maximum level which can be set without the attenuator being adjusted
(Attenuator FIXed).
Return values:
<Upper> float
Default unit: dBm
Example: See Chapter 13.1.2, "Generating an I/Q modulated signal",
on page 215.
Usage: Query only
Manual operation: See "Level Range" on page 102
:OUTPut:AMODe
<AMode>
Switches the mode of the attenuator at the RF output.
Parameters:
<AMode> AUTO | FIXed | APASsive
AUTO
The attenuator is switched automatically. The level settings are
made in the full range.
APASsive
The attenuator is switched automatically. The level settings are
made only for the passive reference circuits. The high-level
ranges are not available.
FIXed
The level settings are made without switching the attenuator.
When this operating mode is switched on, the attenuator is fixed
to its current position and the resulting variation range is defined.
*RST:
AUTO
Example: See Chapter 13.1.2, "Generating an I/Q modulated signal",
on page 215.
Manual operation: See "Mode" on page 101
:OUTPut[:STATe]:PON <Pon>
Selects the state which the RF output assumes when the instrument is switched on.
Parameters:
<Pon> OFF | UNCHanged
*RST: UNCHanged
OUTPut subsystem