[:SOURce]:POWer:ATTenuation:RFOFf:MODE <Mode>
Determines the attenuator's state after the instrument is switched on.
Parameters:
<Mode> MAX | FATTenuated | FIXed | UNCHanged
MAX = FATTenuated
Sets attenuation to maximum when the RF signal is switched off.
This setting is recommended for applications that require a high
level of noise suppression.
FIXed = UNCHanged
Retains the current setting and keeps the output impedance
constant during RF off.
*RST: n.a. (factory preset: MAX)
Manual operation: See "RF-Off-Mode" on page 102
[:SOURce]:POWer:ATTenuation:SOVer[:OFFSet] <Offset>
Sets the switch-over offset value of the attenuator.
Parameters:
<Offset> float
Range: -10 to 10
Increment: 0.1
*RST: 0
Manual operation: See "SATT Switch-Over Offset" on page 102
[:SOURce]:POWer:LMODe <LevMode>
Selects the level mode.
Parameters:
<LevMode> NORMal | LNOise | LDIStortion
NORM
automatic selection of the best settings
LNOISe
settings for lowest noise
LDIStortion
settings for lowest distortions
Example: See Chapter 13.1.4, "Advanced tasks to optimize performance",
on page 218.
Manual operation: See "Mode" on page 100
[:SOURce]:POWer:POWer <Amplitude>
Sets the level at the RF output connector.
SOURce:POWer subsystem