Preparing for use
R&S
®
SGS100A
23User Manual 1173.9105.02 ─ 13
There is no need to set the voltage manually or change fuses.
3.7 Connecting to LAN
Network environment
Before connecting the product to a local area network (LAN), consider the following:
Install the latest firmware to reduce security risks.
For internet or remote access, use secured connections if applicable.
Ensure that the network settings comply with the security policies of your company.
Contact your local system administrator or IT department before connecting your
product to your company LAN.
When connected to the LAN, the product may potentially be accessed from the
internet, which may be a security risk. For example, attackers might misuse or
damage the product.
To connect to LAN
The connector is on the rear panel.
Connect the LAN socket via an RJ-45 cable to the LAN.
By default, the R&S SGS is configured to use DHCP (dynamic host configuration pro-
tocol) and no static IP address is configured.
3.8 Connecting to RF 50Ω
The connector is on the rear panel.
To prepare for connecting to RF 50Ω
1.
NOTICE! Damaged or not clean connections can lead to RF insertion loss and mis-
match, and even premature wear of the connectors.
Before connecting to the port, inspect the RF connector visually to check that it is
clean, undamaged and mechanically compatible.
See the application note 1MA99 for information on how to handle and maintain the
RF port, to minimize measurement deviations and ensure its longevity.
2.
NOTICE! Risk of instrument damage. Excessive reverse power or DC voltage at
the RF 50Ω connector can damage the instrument.
Make sure that the values do not exceed the reverse power and DC limits as given
in the data sheet.
3. If switched-on, deactivate the RF output of the R&S SGS, before connecting an RF
cable to the RF 50Ω connector.
Connecting to RF
50Ω