Remote control commands
R&S
®
SGS100A
294User Manual 1173.9105.02 ─ 13
Parameters:
<Hostname> string
Example:
SYSTem:PROTect1:STATe OFF,123456
SYSTem:COMMunicate:NETWork:HOSTname 'SIGGEN'
sets the individual computer name of the R&S SGS.
Manual operation: See "Hostname" on page 152
:SYSTem:COMMunicate:NETWork[:IPADdress]:GATeway <Gateway>
Sets the IP address of the default gateway.
Parameters:
<Gateway> string
Range: 0.0.0.0 to ff.ff.ff.ff
Example:
SYSTem:COMMunicate:NETWork:IPADdress:GATeway
'1.2.3.4'
sets the IP address of the default gateway.
Manual operation: See "Default Gateway" on page 154
:SYSTem:COMMunicate:NETWork[:IPADdress]:SUBNet:MASK <Mask>
Sets the subnet mask.
Parameters:
<Mask> string
Example:
SYSTem:COMMunicate:NETWork:IPADdress:SUBNet:
MASK '255.255.0.0'
determines the subnet mask.
Manual operation: See "Subnet Mask" on page 153
:SYSTem:COMMunicate:NETWork:RESource?
Queries the VISA resource string, used for remote control of the instrument with VXI-11
protocol.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:NETWork:RESource?
Response: "TCPIP::192.1.2.3::INSTR"
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 154
:SYSTem:COMMunicate:SERial:BAUD <Baud>
Sets the baudrate for the serial remote control interface.
SYSTem subsystem