Example:
PULM:DEL 13 us
13 us elapse after a trigger before the first pulse is generated.
Manual operation: See "Pulse Delay" on page 128
[:SOURce<hw>]:PULM:DOUBle:DELay <Delay>
Sets the delay from the start of the first pulse to the start of the second pulse.
Parameters:
<Delay> float
Range: 40 ns to 100 s
Increment: 10 ns
*RST: 1 ns
Example:
PULM:DOUB:DEL 22 us
22 us elapse between the beginning of the first pulse and the
beginning of the second pulse in double-pulse mode.
Manual operation: See "Double Pulse Delay" on page 128
[:SOURce<hw>]:PULM:DOUBle:WIDTh <Width>
Sets the width of the second pulse in case of double pulse generation.
Example:
PULM:DOUB:WIDT 33 us
sets a width of 33 us for the second pulse.
Manual operation: See "Double Pulse Width" on page 127
[:SOURce<hw>]:PULM:MODE <Mode>
Sets the mode of the pulse generator.
Parameters:
<Mode> SINGle | DOUBle
SINGle
Enables single pulse generation.
DOUBle
Enables double pulse generation. The two pulses are generated
in one pulse period.
*RST: SINGle
Example:
PULM:MODE DOUB
enables double pulse generation.
Manual operation: See "Pulse Mode" on page 127
[:SOURce<hw>]:PULM:PERiod <Period>
Sets the period of the generated pulse. The period determines the repetition frequency
of the internal signal.