Signal generator settings
R&S
®
SGS100A
115User Manual 1173.9105.02 ─ 13
3. Connect/disconnect the R&S NRP sensors
Detection and mapping
The R&S SGS automatically detects a connected R&S NRP power sensor and indi-
cates it in the "NRP Power Viewer" and "NRP Sensor Mapping" dialogs.
You can change the default mapping, see Chapter 9.5.8.2, "NRP sensor mapping",
on page 115.
On connection, the R&S SGS immediately starts the measurement of a detected
R&S NRP power sensor. If you perform an instrument preset, the R&S SGS stops the
measurements. The connection and the mapping of the power sensors remain, the
measurements must be restarted.
9.5.8.2 NRP sensor mapping
The "NRP Sensor Mapping" lists all R&S NRP sensors detected by the instrument.
Any R&S NRP sensor that supports the USB legacy protocol and is connected to one
of the USB interfaces, is detected automatically and added to the list. Vice versa, the
R&S SGS removes a sensor from the list, when it is disconnected.
R&S NRP sensors that are connected via LAN or use the USBTMC protocol are not
automatically detected. They are detected by the scan search function.
Access:
Select "Instrument" > "NRP Sensor Mapping".
Level and power-on settings