Performing configuration tasks
R&S
®
SGS100A
164User Manual 1173.9105.02 ─ 13
The scan function finds out the connected R&S SGU.
5. To enable the R&S SGU as an extension, proceed as follows:
a) Select the R&S SGU from the list
b) Set "Available Instruments > Select > On".
A green status indicator "Connected" indicates the successfully established remote
connection to the extension.
6. Select "Test Signal Connections > Test" to trigger a check of all required signal
connections.
The diagram displays the connection state of the tested connections.
7. Configure RF output signal settings:
a) Select "SGMA-GUI > R&S SGS > Freq = 20 GHz".
b) Select "Lev = -30 dBm".
c) Activate the RF output: Select "RF > State > On".
The extension adopts these values and states automatically. Generated is an I/Q
signal with "RF = 20 GHz" and "Level = -30 dBm".
How to generate I/Q signals