Remote control commands
R&S
®
SGS100A
217User Manual 1173.9105.02 ─ 13
// Response: 1
// Request granted, i.e. the instrument is locked
// Abort program if request is refused
// ******************************************************************
// Query the VISA resource strings
// ******************************************************************
SYSTem:COMMunicate:NETWork:RESource?
// Response: TCPIP::10.113.11.91::INSTR
SYSTem:COMMunicate:SOCKet:RESource?
// Response: TCPIP:rssgs100a100021::5025::SOCKET
SYSTem:COMMunicate:USB:RESource?
// Response: USB::0x0AAD::0x0088::100021::INSTR
SYSTem:COMMunicate:PCIexpress:RESource?
// Response: PCIe::0x0AAD::4909::100021::INSTR
// ******************************************************************
// Query network settings
// ******************************************************************
SYSTem:COMMunicate:NETWork:COMMon:HOSTname?
// Response: rssgs100a100021
SYSTem:COMMunicate:NETWork:IPADdress:MODE?
// Response: AUTO
SYSTem:COMMunicate:NETWork:IPADdress?
// Response: 10.113.11.91
SYSTem:COMMunicate:NETWork:IPADdress:SUBNet:MASK?
//Response: 255.255.0.0
SYSTem:COMMunicate:NETWork:IPADdress:GATeway?
//Response: 10.113.0.1
// ******************************************************************
// Changing network settings
// ******************************************************************
SYSTem:PROTect1:STATe OFF,123456
// SYSTem:COMMunicate:NETWork:COMMon:HOSTname "mySGS"
// SYSTem:COMMunicate:NETWork:IPADdress:MODE STATic
// SYSTem:COMMunicate:NETWork:IPADdress 9.8.7.6
SYSTem:COMMunicate:NETWork:RESTart
// ******************************************************************
// Unlock the instrument
// ******************************************************************
UNL 72349234
Programming examples