:CALibration:IQModulator:FULL?
Starts the adjustment of the I/Q modulator for the entire frequency range. The I/Q mod-
ulator is adjusted with respect to carrier leakage, I/Q imbalance and quadrature.
Return values:
<Modulator> 0 | 1
Example: See Chapter 13.1.1, "Performing general tasks for instrument
setup", on page 213.
Usage: Query only
Manual operation: See " I/Q Modulator" on page 143
:CALibration:IQModulator:LOCal?
Starts the adjustment of the I/Q modulator for the current frequency. The I/Q modulator
is adjusted with respect to carrier leakage, I/Q imbalance and quadrature.
Example:
See Chapter 13.1.1, "Performing general tasks for instrument
setup", on page 213.
Usage: Query only
Manual operation: See "Adjust I/Q Modulator at Current Frequency" on page 142
:CALibration:LEVel[:MEASure]?
Starts all adjustments which affect the level.
Return values:
<Level> 1 | ON | 0 | OFF
Example: See Chapter 13.1.1, "Performing general tasks for instrument
setup", on page 213.
Usage: Query only
Manual operation: See " Level" on page 143
:CALibration:LEVel:TEMPerature?
:CALibration:FREQuency:TEMPerature?
:CALibration:IQModulator:TEMPerature?
Queries the delta temperature since the last performed adjustment.
Return values:
<Temperature> string
Usage: Query only
Manual operation: See "Delta Temperature" on page 143
CALibration subsystem