Remote control commands
R&S
®
SGS100A
271User Manual 1173.9105.02 ─ 13
[:SOURce]:IQ:IMPairment:LEAKage:I <I>
[:SOURce]:IQ:IMPairment:LEAKage:Q <Q>
Sets the carrier leakage amplitude for the I-signal/ Q-signal component.
Parameters:
<Q> float
Range: -5 to 5
Increment: 0.01
*RST: 0
Default unit: PCT
Example: See Chapter 13.1.2, "Generating an I/Q modulated signal",
on page 215.
Manual operation: See "Offset" on page 135
[:SOURce]:IQ:IMPairment:QUADrature[:ANGLe] <Angle>
Sets the quadrature offset for the digital I/Q signal.
Parameters:
<Angle> float
Range: -8 to 8
Increment: 0.01
*RST: 0
Default unit: DEG
Example: See Chapter 13.1.2, "Generating an I/Q modulated signal",
on page 215.
Manual operation: See "Quadrature Offset" on page 135
[:SOURce]:IQ:IMPairment[:STATe] <State>
Activates/ deactivates the three impairment or correction values LEAKage, QUADra-
ture and IQRatio for the baseband signal prior to input into the I/Q modulator.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example: See Chapter 13.1.2, "Generating an I/Q modulated signal",
on page 215.
Manual operation: See "State" on page 134
[:SOURce]:IQ:CREStfactor <CrestFactor>
Sets the crest factor of the IQ modulation signal.
SOURce:IQ subsystem