162
1212
SpecificationsSpecifications
Series Name W Series
Model AP225W AP135W AP125WD AP225WD AP124W AP224W AP324W
Weighing Capacity
220g 135g 120g/52g 220g/102g 120g 220g 320g
Minimum Display 0.01mg 0.01mg 0.1mg/0.01mg 0.1mg/0.01mg 0.1mg 0.1m g 0.1mg
Calibration
Weight
Built-in (*1)
Range of
External Weights
for Calibration
Purposes
(Recommended
weight)
95~
220.00090g
(200g)
45~
135.00090g
(100g)
45~
120.00090g
(100g)
95~
220.00090g
(200g)
45~
120.0090g
(100g)
95~
220.0090g
(200g)
95~
320.0090g
(300g)
Repeatability
(when under
weighing
capacity load)
0.015mg
(up to 20g)/
0.03mg
(up to 100g)/
0.05mg
(up to 200g)
0.05mg
0.1mg /
0.02mg
0.1mg /
0.05mg
0.1mg 0.1m g 0.15m g
Repeatability
(when under low
load) (*2)
0,01mg (load weight 5g) (*3)
0,1mg (load
weight 5g)
0.1mg (load
weight 10g)
0,1mg (load
weight 20g)
Minimum weighing
value (*2)
20mg (*3) 200mg
Linearity ±0.1m g ± 0.1m g
±0.2mg/
±0.05mg
±0.2mg/
±0.1m g
±0.2mg ±0.3mg
Display response
time for small
quantity weighing
(*4*5)
Approx. 2 sec.
Response Time (*5)
Approx. 8 sec.
Approx. 2 sec. / Approx. 8 sec.
Approx. 2 sec.
Temperature/
Humidity Range
During Use
5~40°C, 20~85% (*6)
Sensitivity
Temperature
Coefficient
±2ppm/°C (10~30°C)
Pan Size Approximately 91DIA. mm
Main Unit
Dimensions
Approximately 212(W) × 411(D) × 345(H)mm Approximately 212(W) × 367(D) × 345(H)mm
Weight Approximately 7.9 kg Approximately 7.0 kg
Display Organic EL Display (Dot Matrix)
Rated Power Supply
DC 12V , 1.0A
Degree of Soiling 2
Overvoltage
Category
II
Altitude Maximum 2,000m
Installation
Environment
Limited to Indoor Use
Input Power Supply
(AC Adapter)
AC100-240V, 0.32A 50/60Hz (*7)
I/O Terminals RS232C(D-sub9P Plug) , USB host (Type A) , USB device (Type B), Ionizer
*1 Models with a minimum display of 0.01 mg come standard equipped with 2 built-in weights.
*2 USP41 compliant. Representative values from testing using weights equal to approximately 5% of the weighing capacity (or 5 g) at
the Shimadzu Corporation factory. The minimum display value will be affected by the environment in which the balance is installed,
so measurement needs to be carried out in the actual usage location.
*3 The values are resulted obtained from testing with the draft shield inner plate
*4 Time until 90% of the value of the sample quantity added in small quantity weighing (1 mg) is displayed.
*5 The response time is a representative value.
*6 No condensation.
*7 Depending on the attached AC adapter.
*8 The model names and specifications, etc., listed herein are subject to alteration without prior notice. Access the Shimadzu website
(https://www.shimadzu.com/an/balance/) for the latest information.