EN
DL100 – CANopen
®
Distance measuring device
OPERATING INSTRUCTIONS

Table of Contents

Questions and Answers

  • S
    Sandra MillsAug 27, 2025
    What does 'wTemp' indicate on SICK Measuring Instruments and how can I resolve it?
    • K
      Kathryn TurnerAug 27, 2025
      The error message 'wTemp' on your SICK Measuring Instruments indicates that the internal device temperature is nearing the permissible range. Here are some possible solutions: * Check the ambient temperature and improve ventilation if needed. * Shield the device against radiation heat, such as direct sunlight. * Use a device with heating at low ambient temperatures. * Use cooling housings for high ambient temperatures.
  • R
    Ronald MunozAug 25, 2025
    What does 'wLevel' mean on SICK Measuring Instruments and how to fix it?
    • A
      Albert AdkinsAug 26, 2025
      The error message 'wLevel' on your SICK Measuring Instruments indicates that the current damping value is below the recommended level. To resolve this: * Clean any external lens surfaces like the reflector and the lens. * Decrease the distance between the measuring device and the reflector. * Use a distance measuring device with a higher range.
  • S
    shelby42Aug 23, 2025
    What to do if SICK Measuring Instruments show 'wPlb' error?
    • R
      Robert PooleAug 25, 2025
      If the measured value from your SICK Measuring Instruments is not plausible (wPlb), it's likely because the light path between the measuring device and the reflector is interrupted. Observe the light spot on the reflector to ensure it remains steady. If needed, realign the measuring device and reflector, or consider using a larger reflector. Also, it may be caused by optical interferences. Remove any optical interferences and realign the measuring device and reflector.

Summary

Operating Instructions DL100 - CANopen

Important Safety Notes

Critical safety information for device operation, including laser safety.

General Information

Basic details about the manual, symbols, liability, delivery, and service.

Safety Guidelines

Comprehensive safety rules, intended/non-intended use, and personnel requirements.

Device Identification

Information for identifying the device, including the type label.

Setup and Functionality

Guidance on physical setup, operational principles, display, and controls.

Transport and Storage

Procedures for safe transport, inspection, and storage of the device.

Mounting Instructions

Detailed steps and considerations for installing the distance measuring device.

Electrical Connection

Safety precautions and diagrams for connecting the device electrically.

Operation of the Measuring Device

Instructions for device operation, parameter selection, and value changes.

Operation via Ethernet Interface

Configuration and operation using the Ethernet interface and SOPAS ET.

CANopen Interface

Comprehensive guide to CANopen communication, configuration, and objects.

Cleaning and Maintenance

Procedures for cleaning and routine maintenance of the device.

Troubleshooting Guide

Diagnosing and resolving common issues using LED status, warnings, and errors.

Repair and Technical Data

Information on repair policies and detailed technical specifications.

Device Accessories

Details on available accessories like reflectors, cables, and mounting systems.

Menu Structure

Visual representation of the device's navigation hierarchy and parameters.

SICK DL100 Specifications

General IconGeneral
Wave length650 nm
Resolution0.1 mm
Output signalAnalog
Supply voltage10 V DC ... 30 V DC
Beam Diameter3 mm
Operating Temperature-10 °C ... +50 °C
Storage Temperature-25°C to +70°C
Power SupplyDC