Maintenance and diagnostics
4.2 Diagnostic events
FST020 Function manual
130 Function Manual, 09/2017, A5E42949526-AA
Setting
Auto; Manual
Default
Auto
Path
Maintenance and diagnostics Diagnostic events Acknowledge mode
4.2.5
Transmitter detailed events (read only)
Description
Transmitter Detail Alarms menu is only visible in case an alarm with detailed alarm
information is pending.
Setting
DE10 Failed to read any data related page
DE11 Failed to read FATAL_ERROR_PAGE from
DE12 Failed to read SHUT_DOWN_PAGE from
DE21 The EEPROM data is inconsistent
DE22 Inconsistency of totalizer 1 values
DE26 parameter cache full
DE27 Unsupported sensor units
De28 Communications module not found
DE29 Incomplete sensor connected
DE30 Invalid sensor type
Default Communication module not found
Path
Maintenance and diagnostics Diagnostic events Transmitter detailed events
4.2.6
Suppression time
Description
A coming event can be suppressed in order to prevent alarm flooding.
Setting
0 to 300 s
Default
0 s
Path
Maintenance and diagnostics Diagnostic events Suppression time