Maintenance and diagnostics
4.8 Simulation
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
161
4.8.1.3
Channel 4 - Input/output
4.8.1.3.1 Operation mode
Description
Set the output functionality you want the channel to operate.
Setting
Off; Frequency output; Pulse output
Default
Off
Related
The setting of this parameter affects the setting options in parameter
Simulation value.
Path
Maintenance and diagnostics Simulation Simulation inputs and outputs Channel 4 -
input/output Operation mode
4.8.1.3.2 Simulation
Description
Enables or disables simulation of Channel 4 input/output.
Setting
Disabled; Enabled
Default
Disabled
Related
Only if this parameter is set to Enabled, then the setting of parameter
Simulation value
will
be relevant.
Path
Maintenance and diagnostics Simulation Simulation inputs and outputs Channel 4 -
input/output Simulation
4.8.1.3.3 Simulation value
Description
Set the frequency value to be output on the channel when frequency output simulation is
enabled.
Set the pulse value to be output on the channel when pulse output simulation is enabled.
Setting
0.0 Hz to 12500.0 Hz
Default
1.0 Hz
Operation mode is set to Frequency output