Setup
3.2 Process values
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
51
3.2.5.3
Custom conversion factor
Description
Conversion factor for custom Density values based on SI density units kg/m
3
.
Setting
Default
1.0
Related
Path
Setup Process values Density Custom conversion factor
3.2.5.4
Decimal places
Description
Sets the decimal places for the process value displayed in the local display.
Setting
Auto; No decimal place; 1 decimal place; 2 decimal places; 3 decimal places; 4 decimal
places; 5 decimal places; 6 decimal places
Default
2 decimal places
Related
Path
Setup Process values Density Decimal places
3.2.5.5
Alarm warning limits
3.2.5.5.1 Upper alarm limit
Description
If density value exceeds this limit an alarm is triggered.
Setting
0.00 kg/m³ to +10000.00 kg/m³
Default
900.00 kg/
Related
Path
Setup Process values Density Alarm and warning limits Upper alarm limit
3.2.5.5.2 Upper warning limit
Description
If density value exceeds this limit a warning is triggered.