Setup
3.1 Sensor settings
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
31
Related
Path
Setup Sensor settings Path settings Path 1 Serial number
3.1.7.3
Path diagnostic alarm limits
3.1.7.3.1 Percent bursts accepted alarm limit
Description
Sets the desired limit for the percentage of accepted bursts. If measured percentage
accepted burst drops below this limit for any installed path, an alarm will be issued.
Setting
Min: 0%; Max: 100%
Default
50%
Related
Path
Setup Sensor settings Path settings Path diagnostic alarm limits Percent bursts
accepted alarm limit
3.1.7.3.2 Correlation factor alarm limit
Description
Sets the desired limit for the correlation factor. If the measured correlation factor drops below
this limit for any installed path, an alarm will be issued.
Setting
Min: 0.0; Max: 1.0
Default
0.95
Related
Path
Setup Sensor settings Path settings Path diagnostic alarm limits Correlation factor
alarm limit
3.1.7.3.3 RxGain alarm limit
Description
Sets the limit for the RxGain value. If the RxGain goes above this limit for any installed path,
an alarm will be issued.
Setting
Min: 0.0 dB; Max: 100.0 dB
Default
50.0 dB