EasyManua.ls Logo

Siemens FST020 - Sound Velocity; Density; Kinematic Viscosity

Siemens FST020
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and diagnostics
4.8 Simulation
FST020 Function manual
164 Function Manual, 09/2017, A5E42949526-AA
4.8.2.5
Sound velocity
Simulation value
Description
Set the Sound velocity value to be simulated.
Setting
100.0 m/s to 25000.0
Default
1500.0 m/s
Related
Only if Sound velocity is marked in
Enable simulation
, then the setting of this parameter
will
be relevant.
The range and default setting are dependent on the selected
Unit.
Path
Maintenance and diagnostics Simulation Simulation process values Sound velocity
Simulation value
4.8.2.6
Density
Simulation value
Description
Set the Density value to be simulated.
Setting
0.0 kg/m
3
- 10000.0 kg/m
3
Default
1000.0 kg/m
3
Related
Only if Density is marked in
Enable simulation
, then the setting of this parameter
will be
relevant.
The range and default setting are dependent on the selected
Unit.
Path
Maintenance and diagnostics Simulation Simulation process values Density Simu-
lation value
4.8.2.7
Kinematic viscosity
Simulation value
Description
Set the Kinematic viscosity value to be simulated.

Table of Contents