Setup
3.4 Inputs and outputs
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
75
Setting
Alarm class; Alarm item
Default
Alarm class
Related
Path
Setup Inputs and outputs Channel 3 - relay Status mode
3.4.3.3
Sensor alarms (1)
Description
Selection of alarm items that impact the output. Mainly sensor alarms. Applicable if Status
mode = Alarm item.
Setting
06 DSL Storage; 07 Flow measurement; 08 DSL internal; 14 Path 1: No signal; 26 Sensor
temp. comp.; 28
Default None
Related
Path
Setup Inputs and outputs Channel 3 - relay Sensor alarms (1)
3.4.3.4
Sensor alarms (2)
Description
Selection of alarm items that impact the output. Mainly sensor alarms. Applicable if Status
mode = Alarm item.
Setting
34 Invalid flow meas.; 36 Configuration 2; 37 DSL system monitor
Default
None
Related
Path
Setup Inputs and outputs Channel 3 - relay Sensor alarms (2)
3.4.3.5
Process alarms (1)
Description
Selection of alarm items that impact the output. Applicable if Status mode = Alarm item.